Raytek® Compact CI and CM
Raytek Compact CI
Raytek Compact CI is available in Asia Pacific and Americas only.
The Raytek CI noncontact infrared pyrometer is compact and easy to install. The CI unit’s onboard electronics are protected by a rugged IP 65 (NEMA-4) stainless steel housing, allowing the sensor to function in ambient temperatures to 70°C (160°F) without cooling. Because the CI unit has the same 50 Ohm output impedance as a thermocouple, it functions accurately— without offset errors—when used in conjunction with the thermocouple break protection circuitry in most controllers, displays, and transmitters.
Key Features:
- Read temperature ranges from 0°C to 500°C (32°F to 932°F)
- Withstand ambient temperatures up to 70 °C (160 ºF) without cooling and up to 260°C (500°F) with water cooling
- Compact, rugged steel housing with IP65 (NEMA-4) rating
- Accessories for cooling and air purging
- Powered by 12–24 VDC at 20 mA
- Response time: 350 mSec (95%)
- 4:1 optics at 90% energy
Raytek Compact CM
Although small in size, the Compact CM is loaded with advanced features set for temperature monitoring in OEM applications. These include a 24 volt DC alarm output triggered by target temperature or head ambient temperature, peak hold, valley hold or variable averaging signal processing, adjustable target emissivity settings and adjustable window transmissivity setting. An RS232 digital interface allows the user to configure all programmable sensor variables with the powerful DataTemp® Multidrop software included with each sensor. The built-in sensor health LED provides a convenient on-line indicator of the sensor’s operating status and aids in troubleshooting initial sensor set up.
Key Features:
- Read temperature ranges from -20 to 500 °C (-4 to 932 °F)
- Withstand ambient temperatures up to 70 °C (160 ºF) without cooling and up to 260°C (500°F) with water cooling
- Compact, rugged steel housing with IP65 (NEMA-4) rating
- Accessories for air purging and lens protection
- Powered by 12–24 VDC at 20 mA
- Response time: 150 mSec (95%)
- 13:1 optics at 90% energy
- Advanced feature set for OEM applications
- DataTemp Multidrop software compatible
- Sensor status/self diagnostic LED indicator
| Model Number | CI-A | CI-B | CM |
|---|---|---|---|
| Temperature Range | 0 to 350 °C (32 to 662 ºF) |
30 to 500 ºC (86 to 932 ºF) |
-20 to 500 ºC (-4 to 932 ºF) |
| Spectral Response | 7 to 18 μm | 7 to 18 μm | 8 to 14 μm |
| System Accuracy | 0ºC to 115ºC (32ºF to 240ºF) ±3ºC (±6ºF) 116ºC to 350ºC (241ºF to 662ºF) >±5% |
30ºC to 99ºC (86ºF to 211ºF) ±6ºC (±10ºF) 100ºC to 500ºC (212ºF to 932ºF) >±2% or ±3ºC (±6ºF) |
Digital/voltage: 1.5% of reading or ± 2ºC* T/c: ± 1.5% of reading ± 2ºC or ± 4ºC* |
| Repeatability | ±1% of measured value or ±1°C (2°F)* | Digital/voltage: ± 0.5% of reading or ± 1ºC* T/c: ± 0.5% of reading ± 1ºC or ± 2ºC* |
|
| Response Time (90% of time) | 350 ms | 150 ms | |
| Optical Resolution (D:S) | 4 : 1 | 13 : 1 | |
| Signal Processing | – | Peak hold, valley hold, variable averaging | |
| Emissivity | Fixed at 0.95 | 0.10 to 1.10 (adjustable) | |
| Self diagnostics/sensor status | – | Normal operation, alarm, out of range, unstable, sensor fault | |
| Compliance | IEC 801-4, Level 1 | EC – Directive 2004/108/EC (EMC) | |
Compact CI and CM pyrometers can be used across numerous industries and are ideal for a variety of OEM and end-user manufacturing applications including the list below and many others.
Click here if you would like to speak with an expert to discuss your application
- Paper processing
- Plastic Processing
- Asphalt
- Textile Manufacturing
Each Compact CI and CM sensor includes two mounting nuts, 1m (3.2ft.) of cable, and an operator’s manual. Longer cables up to 15m (50ft.) maximum are available and must be specified at the time of order.
Accessories for Compact CI Pyrometers
|
Accessories for Compact CM Pyrometers
|